Autoclave Unloading Isolator

It is used to connect to the sterilization cabinet. It provides an aseptic environment for the assembly of sterilized articles and prevents microbial contamination.

    ? The equipment cavity is made of SUS316L stainless steel with a thickness of 3 mm, and the visual window is made of 15 mm tempered glass.

    ? The system conforms to the ergonomic design, cleanliness is ISO 4.8, and a full set of 4q verification services is available.

    ? The isolator is integrated with the temperature and humidity monitoring system, differential pressure monitoring system, air speed monitoring system, hydrogen peroxide concentration monitoring system, and the particle and plankton monitoring system.

    ? The isolator can detect the leak rate online (the pressure attenuation method), which meets the data requirements of ISO 10648, ISO 14644 and PDA TR34.

    ? The system design meets the GMP, ISO, ISPE, pharmacopoeia and other relevant requirements and complies with 21CFR Part11, and the product has passed the EU CE certification.

    ? RTP valves can be configured according to process requirements.

    Cleanliness level

    Leak rate

    Sterility assurance level

    Noise

    Illuminance

    Differential pressure range

    Fissure air speed

    Grade A

    1%

    SAL≤10-6

    ≤65dB

    ≥500lux

    -0pa~+ 80Pa

    ≥0.5m/s